PolyTrans P 32000 (KNICK)科尼克模块 输入:热电阻,热电偶信号.输出:0 ... 20 mA,4 ... 20 mA,0 ... 5 V,0 ... 10 V.测试电压:2.5KV.电源:24VDC. | Input data | Type of sensor | Standard | Measuring range | | | | | | Input 1) | Pt100 | DIN 60751 | -200 ... +850 °C | | | Pt1000 | DIN 60751 | -200 ... +850 °C | | | Other platinum resistors | DIN 60751 | -200 ... +850 °C | | | Ni100 | DIN 43760 | -60 ... +180 °C | | | Other nickel resistors | DIN 43760 | -60 ... +180 °C | | | | Connection | 2-, 3- or 4-wire (automatically recognized) | | | signaled by a yellow LED | | | | Range of resistance incl. cable resistance | for temperature detection: 5 kohms | | | for resistance detection: 0 ... 5 kohms or 5 ... 100 kohms | | | | Max. cable resistance | 100 ohms | | | | Supply current | 200 µA, 400 µA or 0 ... 500 µA | | | | Wire monitoring | Open circuit detection | | | | Input error limits | Resistance <5 kohms: | ± (50 mohms + 0.05 % meas. val.) for ranges > 15 ohms | | | Resistance >5 kohms: | ± (1 ohm + 0.05 % meas. val.) for ranges > 50 ohms | | | | Input temperature coefficient | <50 ppm/K of selected end value | | | (medium temperature coefficient in the allowed temperature range, reference temperature 23 °C) |
| Thermocouples | | | | | Input data | Type of sensor | Standard | Measuring range | | | | | Input 2) | Typ B | DIN 60584-1 | +250 … +1820 °C | | | Type E | DIN 60584-1 | –200 … +1000 °C | | | Type J | DIN 60584-1 | –210 … +1200 °C | | | Type K | DIN 60584-1 | –200 … +1372 °C | | | Type L | DIN 43710 | –200 … +900 °C | | | Type N | DIN 60584-1 | –200 … +1300 °C | | | Type R | DIN 60584-1 | –50 … +1767 °C | | | Type S | DIN 60584-1 | –50 … +1767 °C | | | Type T | DIN 60584-1 | –200 … +400 °C | | | Type U | DIN 43710 | –200 … +600 °C | | | W3Re/W25Re | ASTM E988-96 | 0 … +2315 °C | | | W5Re/W26Re | ASTM E988-96 | 0 … +2315 °C | | | | Input resistance | >10 Mohms | | | | Max. cable resistance | 1 kohms | | | | Wire monitoring | Open circuit detection | | | | Input error limits | ± (10 µV + 0.05 % meas. val.) for spans >2 mV | | | | Input temperature coefficient | <50 ppm/K of selected end value | | | (medium temperature coefficient in the allowed temperature range, reference temperatur 23 °C) | | | | Cold junction compensation | Internal (Pt100) | | | Selectable via IrDA®: external (Pt100), fixed value or uncompensated | | | | Error of internal cold junction compensation | <1.5 K | | | | Error of external cold junction compensation | < 80 mohms + 0.1 % meas. val. | by Pt100 for Tcomp= 0 ... 80 °C |
| Shunt voltages | | | | Input data | | | | Input | -1000 ... 1000 mV unipolar/bipolar | | | | Input resistance | >10 Mohms | | | | Input error limits | ± (200 µV + 0.05 % meas. val.) for spans >50 mV | | | | Wire monitoring | Open circuit detection | | | | Input temperature coefficient | <50 ppm/K of selected end value | | | (medium temperature coefficient in the allowed temperature range, reference temperature 23 °C) | | | | Overload | 5 V across all inputs |
| Input data Strain gages | | | | Input | ± 7.5 mV/V | | | | Bridge resistance | 200 ohms ...10 kohms | | | | Zero adjustment | Within input range | | | | Supply current (internal supply) | 0 ... 5 mA | | | | Supply voltage (external supply) | 1 ... 3 V | | | | Input error limits | ± (2 µV/V + 0.1 % meas. val.) for spans ≥0.5 mV/V | | | | Wire monitoring | Open circuit and short circuit detection | | | | Input temperature coefficient | <50 ppm/K of selected sensitivity | | | (medium temperature coefficient in the allowed temperature range, reference temperature 23 °C) | | | | Overload | 5 V across all inputs |
| Potentiometers | | | Input data | | | | Input | 200 ohms ... 50 kohms | | | | Connection | 3- or 4-wire | | | | Feeding current | 0 ... 5 mA | | | | Wire monitoring | Open circuit and short circuit detection | | | | Input error limits | ± (0.2 % full scale + 0.05% meas. val.) for spans >5 % | | | | Input temperature coefficient | <50 ppm/K of selected end value | | | (medium temperature coefficient in the allowed temperature range, reference temperature 23 °C) |
| Output data | | | | Outputs | 0 ... 20 mA, | calibrated switching | | | 4 ... 20 mA, | (default 4 ... 20 mA) | | | 0 ... 5 V, | | | 0 ... 10 V | | | | Control range | 0 ... approx. 102,5 % span at 0 ... 20 mA, 0 ... 10V and 0 ... 5 V outputs | | | -1.25 ... approx. 102.5 % span at 4 ... 20 mA output | | | | Resolution | 16 bit | | | | | Simulation mode | 0 ... 20 mA current output: | 0 ... 21 mA | | adjustable via IrDA® | 4 ... 20 mA current output: | 3 ... 21 mA | | | 0 ... 5 V voltage output: | 0 ... 5.25 V | | | 0 ... 10 V voltage output: | 0 ... 10.5 V | | | | | Load | Current output: | ≤10 V (≤500 ohms at 20 mA) | | | Voltage output: | ≤1 mA (≥10 kohms at 10 V) | | | | | Output error limits | Current output: | ± (10 µA + 0.05 % meas. val.) | | | Voltage output: | ± (5 mV + 0.05 % meas. val.) | | | | | Residual ripple | <10 mVrms | | | | Output temperature coefficient | <50 ppm/K of selected end value | | | (medium temperature coefficient in the allowed temperature range, reference temperature 23 °C) | | | | Error signaling | 0 ... 20 mA Output: I =0 mA or ≥21 mA | | | 4 ... 20 mA Output: I ≤3.6 mA or ≥21 mA | | | 0 ... 5 V and 0 ... 10 V Output: V= 0 V or V≥5.25 V and V≥10.5 V respectively | | | by output signal, red LED and IrDA® for | | | out-of-range conditions, | | | incorrect parameter setting, | | | sensor short-circuit and open circuit | | | output error load | | | unintentional adjustment of switches in service (SIL devices) | | | other device errors. See also "Error signaling". |
| Transmission behavior | | | | Characteristic | Linear rising / falling; | | | characteristics definable by interpolation points via IrDA® interface | | | | Measuring rate | Approx. 3 / s *) |
| Display | | | | Green LED | Auxiliary power | | | | Yellow LED | Type of connection | | | IrDA® communication | | | | Red LED | Maintenance request or device breakdown | | | | Power supply | 24V DC power supply | Broad-range power supply | | | | | Power supply | 24 V DC (-20 %, +25 %), approx. 1,2 W | 24 V ... 110 V DC (±20 %), approx. 1.2 W | | | Power supply can be routed from one device to another | 110 V ... 230 V AC (±10 %), 48 ... 62 Hz, approx. 1.5 VA | | | via DIN rail bus-connectors. |
| Isolation | | | | Galvanic isolation | 3-port isolation between input, output and power supply | | | | Test voltage | 2.5 kV AC, 50 Hz: input against output against power supply | | | | Working voltage | Up to 300 V AC/DC across all circuits at overvoltage category II and pollution degree 2 to EN 61010-1 | | (basic insulation) | For applications with high working voltages, you should ensure there is sufficient spacing or isolation from neighboring devices and protection against electric shocks. | | | | Protection against electric | Safe Isolation to EN 61140 by reinforced insulation to EN 61010-1. | | shock | Working voltage across all circuits up to 300 V AC/DC at overvoltage category II and pollution degree 2. | | | For applications with high working voltages, you should ensure there is sufficient spacing or isolation from neighboring devices and protection against electric shocks. |
| Standards and approvals | | | | Functional safety | SIL 2 to IEC 61508, up to SIL 3 when assembled redundantly | | | | Explosion protection | ATEX Zone 2 (EN 60079-15) | | | Class 1, Div 2 / Zone 2 (UL 1604) | | | | | EMC | Product standard: | EN 61326 | | | Emitted interference: | Class B | | | Immunity to interference 3): | Industry | | | EMC requirements for devices with safety related functions IEC 61326-3 | | | | cURus | File No. 220033 | | | Standards: UL 508 and CAN/CSA 22.2 No. 14-95 |
| Interfaces | | | | IrDA® | Specification 1.1, Slave device for bi-directional communication | | | Communication software Paraly® SW 111 | | | free download at www.knick.de |
| Other data | | | | Ambient temperature | Operation: | 0 ... +55 °C when placed without spacing | | | | 0 ... +65 °C with spacings ≥ 6 mm | | | Transport and storage: | -25 ... +85 °C | | | | Ambient conditions | Stationary use, weatherproof | | | Relative humidity: 5 ... 95 %, no condensation | | | Air pressure: 70 ... 106 kPa | | | Water or wind-driven precipitation (rain, snow, hail etc.) excluded | | | | Design | Modular housing with screw terminals, 6.2 mm wide | | | for further measurements and mounting profile see "Dimension drawings" | | | | Protection class | Terminals IP20, Housing IP40 | | | | Mounting | For 35 mm top-hat rail to EN 60715 | | | | Weight | Approx. 60 g | | | | | | 1) other sensor types with resistance values up to max. 5 kohms on request | | 2) other thermocouples types on request | | *) for thermocouples with external cold junction compensation and for 5 ... 100 kohms resistance detection: approx. 2 / s | | 3) slight deviations are possible while there is interference |